Submitted by Professor David R Dance & Professor Kenneth C Young, NCCPM
   Commenting on behalf of the organisation
Document Diagnostic Reference Levels in Medical Imaging
 
Paragraph 153: States that:

DG can be calculated from Ka,e measured at the surface of the phantom with a suitable calibrated detector using standard equations and conversion factors (Dance et al., 2000; IPEM, 2005).


Please note that the standard equations and conversion factors are for Ka,i  and not for Ka,e. To use Ka,e backscatter factors would be required. These are not standard and are not provided in the references cited.


The references provided for the conversion factors are incomplete. Please consider adding the following:-


Dance D R 1990 Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Phys. Med. Biol. 35 1211-1219


Dance D R, Young K C and van Engen R E 2009  Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA dosimetry protocols. Phys. Med. Biol  54 4361-72

Dance D R, Young K C and van Engen R E 2011  Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA dosimetry protocols. Phys. Med. Biol  56 453-471

Dance D R, Young K C 2014 Estimation of mean glandular dose for contrast enhanced digital mammography: factors for use with the UK, European and IAEA breast dosimetry protocols. Phys. Med. Biol  59 2127-2137